https://elar.uspu.ru/handle/ru-uspu/56248| Title: | Gamma-Densitometer for Studies of High-Temperature Metal Melts |
| Authors: | Rusanov, B. A. Baglasova, E. S. Popel, P. S. Sidorov, V. E. Sabirzyanov, A. A. |
| Issue Date: | 2018 |
| Publisher: | Pleiades Publishing |
| metadata.dc.rights: | info:eu-repo/semantics/openAccess |
| Abstract: | The set-up for the measurement of the density of high-temperature melts by means of penetrating gamma-radiation is described. The modern device base, which makes it possible to automatize the work, is the main difference from the earlier existing gamma-densitometers. The key characteristics of the absolute measurements with the use of very exact values of the mass attenuation coefficient for the gamma-quanta beam are described and the accuracy of measurements is estimated. The results of the density measurement of the molten aluminum, cadmium, tin, and bismuth are presented; they confirm the a priori estimate of the error. © 2018, Pleiades Publishing, Ltd. |
| Keywords: | METEOROLOGICAL INSTRUMENTS A-PRIORI ESTIMATES ABSOLUTE MEASUREMENTS ACCURACY OF MEASUREMENTS GAMMA-QUANTA BEAMS HIGH TEMPERATURE HIGH TEMPERATURE MELTS KEY CHARACTERISTICS MASS ATTENUATION COEFFICIENTS DENSITOMETERS |
| ISSN: | 0018151X |
| DOI: | 10.1134/S0018151X18020190 |
| Appears in Collections: | Научные публикации, проиндексированные в SCOPUS и WoS |